DOI 10.1080/10255842.2026.2696324
Effects of CFR-PEEK plate layup and screw configuration on tibial shaft fracture healing: a simulation study based on a mechanobiological model.
Аннотация
Based on the biphasic mechanoregulation theory, this study developed a mechanobiological model integrates biomechanical stimulus calculation, cellular behaviors, and dynamic updating of callus tissue modulus, to simulate the healing process of tibial shaft fractures stabilized with CFR-PEEK bone plates. The effects of three layup schemes and five screw configurations were evaluated. The model predictions indicate that the combination of the [±45]4S layup with the flexible C5 screw configuration provides appropriate mechanical stimulation during the mid-to-late healing phase, effectively promotes endochondral ossification, and yields the most favorable simulated healing outcome.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.