Preliminary Application of a Continuous Functional Contrast Visual Acuity System in the Assessment of Visual Function in Dry Eye Patients.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Shi Gui-Lian, Pan An-Peng, Hu Rui-Lin, Zhang Yu-Qian, Ma Yun-Jing, Yu A-Yong

Аннотация

PURPOSE: To investigate the feasibility and efficacy of a continuous functional contrast visual acuity (CFCVA) system in the assessment of visual function in dry eye disease (DED).

METHODS: Twenty patients with DED and 15 normal controls were recruited. Subjective symptoms were evaluated using the Ocular Surface Disease Index (OSDI) questionnaire, and tear film stability was assessed by a noninvasive corneal topographer. Under natural blinking conditions, the custom-built CFCVA system was used to take serial visual acuity measurements at 100%, 25%, 10%, and 5% contrast for 60 seconds. A 5-minute measurement at a 100% contrast level was defined as the stress test (ST). Mean CFCVA was defined, and visual maintenance ratio (VMR) was the ratio of mean CFCVA divided by baseline visual acuity.

RESULTS: In both groups, VMR decreased and mean CFCVA (logarithm of the minimum angle of resolution) increased with decreasing optotype contrast (from 100% to 5%). In ST, the ST VMR at the fourth and fifth minutes (VMR54 and VMR55) showed the strongest correlations with OSDI total, ocular symptoms, and vision-related function (-0.646 and -0.598, -0.688 and -0.693, and -0.599 and -0.555, respectively, P < 0.05). VMR54 and VMR55 also demonstrated the best discriminating ability for detecting DED, with areas under the curve of 0.903 and 0.867, respectively.

CONCLUSIONS: Extending the continuous measuring time was more effective for detecting vision-related functional abnormalities in patients with DED than simply decreasing the optotype contrast level.

TRANSLATIONAL RELEVANCE: The proposed CFCVA system and associated parameters offer a potential method for quantifying and interpreting the visual symptoms of DED in clinical care.