An iOS-based VoiceScreen application: feasibility for use in clinical settings-a pilot study.

Uloza Virgilijus, Ulozaite-Staniene Nora, Petrauskas Tadas

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To elaborate the application suitable for smartphones for estimation of Acoustic Voice Quality Index (AVQI) and evaluate its usability in the clinical setting.

METHODS

An elaborated AVQI automatization and background noise monitoring functions were implemented into a mobile "VoiceScreen" application running the iOS operating system. A study group consisted of 103 adult individuals with normal voices (n = 30) and 73 patients with pathological voices. Voice recordings were performed in the clinical setting with "VoiceScreen" app using iPhone 8 microphones. Voices of 30 patients were recorded before and 1 month after phonosurgical intervention. To evaluate the diagnostic accuracy differentiating normal and pathological voice, the receiver-operating characteristic statistics, i.e., area under the curve (AUC), sensitivity and specificity, and correct classification rate (CCR) were used.

RESULTS

A high level of precision of AVQI in discriminating between normal and dysphonic voices was yielded with corresponding AUC = 0.937. The AVQI cutoff score of 3.4 demonstrated a sensitivity of 86.3% and specificity of 95.6% with a CCR of 89.2%. The preoperative mean value of the AVQI [6.01(SD 2.39)] in the post-phonosurgical follow-up group decreased to 2.00 (SD 1.08). No statistically significant differences (p = 0.216) between AVQI measurements in a normal voice and 1-month follow-up after phonosurgery groups were revealed.

CONCLUSIONS

The "VoiceScreen" app represents an accurate and robust tool for voice quality measurement and demonstrates the potential to be used in clinical settings as a sensitive measure of voice changes across phonosurgical treatment outcomes.

Для цитирования: Uloza Virgilijus, Ulozaite-Staniene Nora, Petrauskas Tadas. An iOS-based VoiceScreen application: feasibility for use in clinical settings-a pilot study.. European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery. 2022. DOI: 10.1007/s00405-022-07546-w.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.