Endovascular aortic arch repair with a pre-cannulated double-fenestrated physician-modified stent graft: a benchtop experiment.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Lounes Youcef, Chassin-Trubert Lucien, Gandet Thomas, Ozdemir Baris Ata, Peyron Antoine, Akodad Mariama, Alric Pierre, Canaud Ludovic

Аннотация

OBJECTIVES: The critical step in total endovascular aortic arch repair is to ensure alignment of fenestrations with, and thus maintenance of flow to, supra-aortic trunks. This experimental study evaluates the feasibility and accuracy of a double-fenestrated physician-modified endovascular graft [single common large fenestration for the brachiocephalic trunk and left common carotid artery and a distal small fenestration for left subclavian artery (LSA) with a preloaded guidewire for the LSA] for total endovascular aortic arch repair.

METHODS: Eight fresh human cadaveric thoracic aortas were harvested. Thoracic endografts with a physician-modified double fenestration were deployed for total endovascular aortic arch repair in a bench test model. A guidewire was preloaded through the distal fenestration for the LSA. All experiments were undertaken in a hybrid room under fluoroscopic guidance with subsequent angioscopy and open evaluation for assessment.

RESULTS: Mean aortic diameter in zone 0 was 31.3 ± 3.33 mm. Mean duration for stent graft modification was 20.1 ± 5.8 min. Mean duration of the procedure was 24 ± 8.6 min. The Medtronic Valiant Captivia stent graft was used in 6 and the Cook Alpha Zenith thoracic stent graft in 2 cases. LSA catheterization was technically successful with supra-aortic trunk patency in 100% of cases.

CONCLUSIONS: The use of a double-fenestrated stent graft with a preloaded guidewire appears to be a useful technical addition to facilitate easy and correct alignment of stent graft fenestrations with supra-aortic trunk origins.