DOI 10.1002/jcla.23671

Медицинская библиотека · Journal of clinical laboratory analysis · 2020

Performance evaluation of immunoassay for infectious diseases on the Alinity i system.

Nam Minjeong, Song Da Young, Song Sang Hoon, Roh Eun Youn, Shin Sue, Park Kyoung Un, Song Eun Young

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Although a diagnosis of infectious diseases is essential for timely treatment, the performance of diagnostic tests has been hardly evaluated due to variable results that are influenced by multiple factors in different conditions. In the present study, the performance of the Alinity i system, which is a newly developed immunoassay to diagnose infectious diseases, was evaluated.

METHODS

We evaluated the precision, linearity, correlation, and carryover of 16 analytes (HAV Ab IgG, HBsAg, HBeAg, anti-HBc, anti-HBe, anti-HBs, anti-HCV, HIV Ag/Ab, EBV VCA IgM, EBV VCA IgG, EBV EBNA IgG, CMV IgM, CMV IgG, Toxoplasma IgG, Rubella IgG, and Syphilis TP) of Alinity i by comparison with ARCHITECT i2000SR system following the rationale of the Clinical and Laboratory Standards Institute (CLSI).

RESULTS

For quantitative tests, the coefficients of variation (CV) % of repeatability and intermediate precision were between 0% and 4.18%. The coefficients of the linearity (r2 ) over a widely tested analytical range were ≥ 0.990 and the correlation between Alinity i and the ARCHITECT i2000SR system was strong (r ≥ 0.994). For qualitative tests, the agreement between Alinity i and the ARCHITECT i2000SR system was excellent (kappa coefficient 1) with 100% sensitivity and specificity. Carryover rates for all analytes were less than 1.0% (-0.11% ~ 0.21%).

CONCLUSION

The Alinity i system showed good analytical performance and favorable comparability with the ARCHITECT i2000SR. It could be suitable as a routine immunoassay analyzer for screening and diagnosis of infectious disease.

Для цитирования: Nam Minjeong, Song Da Young, Song Sang Hoon, Roh Eun Youn, Shin Sue, Park Kyoung Un, Song Eun Young. Performance evaluation of immunoassay for infectious diseases on the Alinity i system.. Journal of clinical laboratory analysis. 2020. DOI: 10.1002/jcla.23671.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.