DOI 10.1038/jp.2017.94

Медицинская библиотека · Journal of perinatology : official journal of the California Perinatal Association · 2017

Comparison between Bilistick System and transcutaneous bilirubin in assessing total bilirubin serum concentration in jaundiced newborns.

Greco C, Iskander I F, Akmal D M, El Houchi S Z, Khairy D A, Bedogni G, Wennberg R P, Tiribelli C, Coda Zabetta C D

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To compare the performance and accuracy of the JM-103 transcutaneous bilirubinometer and Bilistick System in measuring total serum bilirubin for the early identification of neonatal hyperbilirubinemia.

STUDY

DESIGN

The study was performed on 126 consecutive term and near-term (⩾36 weeks' gestational age) jaundiced newborns in Cairo University Children Hospital NICU, Egypt. Total serum bilirubin was assayed concurrently by the clinical laboratory and Bilistick System and estimated using the JM-103 transcutaneous bilirubin instrument. Bland-Altman analysis was used to evaluate the agreement between determinations.

RESULT: The limits of agreement of the Bilistick System (-5.8 to 3.3 mg dl-1) and JM-103 system (-5.4 to 6.0 mg dl-1) versus the clinical laboratory results were similar.

CONCLUSION

The Bilistick System is an accurate alternative to transcutaneous (TcB) determination for early diagnosis and proper management of the neonatal jaundice.

Для цитирования: Greco C, Iskander I F, Akmal D M, El Houchi S Z, Khairy D A, Bedogni G, Wennberg R P, Tiribelli C, Coda Zabetta C D. Comparison between Bilistick System and transcutaneous bilirubin in assessing total bilirubin serum concentration in jaundiced newborns.. Journal of perinatology : official journal of the California Perinatal Association. 2017. DOI: 10.1038/jp.2017.94.