A comparison between a hinged plate and screw system and a conventional tension-band plate and screw system used for correction of an angular deformity of the lower limb: an animal study.
Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Wu Zhenkai, Zhao Dahang, Zhao Li, Liu Jianlin, Li Hai, Zhu Jie, Ma Fengcang, Porter Daniel Edward
Аннотация
BACKGROUND: The purpose of the animal study is to introduce a newly designed hinged plate and screw system for correction of angular deformities of the lower limbs. The technique was compared with the use of a conventional tension-band plate and screw system.
METHODS: This is a randomized controlled animal trial. Eight 3-month-old Bama miniature pigs were used to establish animal models. In each animal, one leg was randomly allocated into study group and another leg into control group. Legs of the study group were corrected with a hinged plate and screw system, and legs of the control group were corrected with a conventional tension-band plate and screw system. The corrective rates of medial slope angle, medial proximal tibial angle, and angle of the two arms of the hinged plate were measured. Residual stress on the implants was also evaluated. A P < 0.05 was statistical significant.
RESULTS: At the final measurements of 18 weeks, the mean corrective rates of medial slope angle, medial proximal tibial angle, and angle of the two arms of the study group were 0.71°/week, 0.85°/week, and 2.18°/week, respectively; the data in the control group were 0.84°/week, 0.89°/week, and 2.13°/week, respectively. No significant difference was found between the groups regarding the mean corrective rates of the angles (P < 0.05). The mean residual stress in the study group was 643.35 MPa, and measurement in the control group was 1,273.63 MPa, with a significant difference (P < 0.05).
CONCLUSIONS: Compared to the conventional tension-band plate and screw system, the hinged plate and screw system may be more reliable for correction of angular deformities of the lower limb.
METHODS: This is a randomized controlled animal trial. Eight 3-month-old Bama miniature pigs were used to establish animal models. In each animal, one leg was randomly allocated into study group and another leg into control group. Legs of the study group were corrected with a hinged plate and screw system, and legs of the control group were corrected with a conventional tension-band plate and screw system. The corrective rates of medial slope angle, medial proximal tibial angle, and angle of the two arms of the hinged plate were measured. Residual stress on the implants was also evaluated. A P < 0.05 was statistical significant.
RESULTS: At the final measurements of 18 weeks, the mean corrective rates of medial slope angle, medial proximal tibial angle, and angle of the two arms of the study group were 0.71°/week, 0.85°/week, and 2.18°/week, respectively; the data in the control group were 0.84°/week, 0.89°/week, and 2.13°/week, respectively. No significant difference was found between the groups regarding the mean corrective rates of the angles (P < 0.05). The mean residual stress in the study group was 643.35 MPa, and measurement in the control group was 1,273.63 MPa, with a significant difference (P < 0.05).
CONCLUSIONS: Compared to the conventional tension-band plate and screw system, the hinged plate and screw system may be more reliable for correction of angular deformities of the lower limb.