Heating of Newborn Infants due to Blue Light-Emitting Diode Fibreoptic Phototherapy Pads.

СтатьяНаблюдательное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ng Pei Ling, Carlisle Tony, Ly Marleesa, Morris Scott Adam

Аннотация

BACKGROUND: Surface temperatures of fibreoptic phototherapy pads using a high intensity blue light-emitting diode (LED) light source have not been studied.

OBJECTIVES: The aim of this study was to measure the temperature of LED fibreoptic phototherapy pads during phototherapy in a bench-top study, and to determine temperature effects on babies during phototherapy.

METHODS: A commercially available LED fibreoptic phototherapy system was tested. In a bench-top setting, pad surface temperatures were measured before, during and after a 12-h period of phototherapy (10 different LED light box-pad combinations). A prospective, cohort study of well babies at >34 weeks' gestation receiving phototherapy was then conducted to determine changes in pad and body temperatures during a 90-min phototherapy period.

RESULTS: In the bench-top study, the mean (95% CI) pad temperature was 21.8°C (21.5-22.1) before lights, 27.0°C (26.5-27.5) after 12 h of lights, and 22.1°C (21.9-22.4) 8 h after turning off the lights (F = 366.1, p < 0.0005). The magnitude of change in pad temperature with phototherapy was linearly correlated with irradiance (r = 0.89, p < 0.0005). The pad plastic covering absorbed 13% of blue light from fibres. In the clinical study, the warmest pad temperature during phototherapy was 38.9°C. Axillary temperature increased by a mean (95% CI) of 0.3°C (0.1-0.5), p < 0.019, and exceeded 37.5°C in 4 babies.

CONCLUSIONS: LED fibreoptic phototherapy pads are heated by high-intensity blue light. The thermal environment and temperature of babies should be monitored closely during LED fibreoptic phototherapy. A temperature probe placed between the skin and the pad will not accurately reflect the core temperature during fibreoptic phototherapy.