Short- and long-term histopathologic evaluation of stenting using a self-expanding nitinol stent in pig carotid and iliac arteries.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Verheye S, Salame M Y, Robinson K A, Post M J, Carrozza J P, Baim D S, Sigwart U, King S B, Chronos N A
Аннотация
Stenting is increasingly being used to treat carotid artery disease. However, complications including distal embolization, stent thrombosis, stent collapse from external compression, the need for high-pressure inflation with increased neointimal response, or balloon rupture during stent expansion and stent loss are all potential problems and of concern. To address each of these specific concerns, a new stent was designed, which is self-expandable, made of nitinol, with temperature-dependent superelastic properties, and with high vessel wall surface coverage. Since this device has a number of novel characteristics, we aimed to assess the short- and long-term histopathologic response in pig carotid and iliac arteries. Single stents were deployed in pig carotid and iliac arteries after overstretch balloon injury. Angiograms were performed pre- and poststenting and prior to sacrifice. Intravascular ultrasound was used before implantation to determine vessel size. Vessels were examined histologically at 1 month (n = 6) and 6 months (n = 6) for morphometric analysis, hemorrhage and thrombus, endothelialization, and inflammatory and fibrotic responses. There was a 100% angiographic success rate at implantation. In one case, it was determined histologically that a single stent was implanted in a dissection plane of a pig's left iliac artery and was occluded by organized thrombus, with the true lumen being patent. At 6-month follow-up, this was the only evidence of a single stent occlusion, with flow adjacent to the stent in the true lumen. In the other vessels, the stents showed good vessel wall-stent apposition and the lumens were patent with a concentric and thin neointima. Inflammatory cells were rare and there were no mural thrombi. Coverage of the vessel wall by endothelial-like cells was complete at 1 month. The novel nitinol EndoStent appears to have favorable biocompatibility with minimal thrombus deposition or inflammatory response, and its use is feasible for clinical application in carotid and iliac arteries.