Applied biosafety : journal of the American Biological Safety Association · 2025
A Novel Method to Sterilize Isolators for the Housing of Germ-Free Birds and Production of Germ-Free Eggs.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Murphy Martin, Jones Rhys, Dos Santos Ariane, Adeyemi Modupeore, Caulfield Catherine
Аннотация
INTRODUCTION: The effective sterilization of isolators prior to placement of germ-free birds/eggs is crucial to ensuring that the environment is free of potential contaminants. With the use of formaldehyde for sterilization becoming less popular owing to its carcinogenicity, the need for an alternative agent with the same efficacy is essential in the preparation of isolators. Chlorine dioxide dry gas has previously been shown to be a highly effective sterilization method, providing a promising alternative for germ-free avian egg facilities.
METHODS: Polyvinyl chloride (PVC; n = 32) and stainless-steel production isolators (n = 7) were sterilized using approximately 4 h of exposure to chlorine dioxide (PVC isolator) and after approximately 6 h of exposure to chlorine dioxide (stainless-steel production isolator).
RESULTS: Each isolator type was sterilized effectively using chlorine dioxide. Samples collected for microbiological analysis from the isolators after sterilization confirmed that the isolators were sterilized and remained sterile for at least 3 weeks after sterilization.
DISCUSSION: The results of this study highlight the first use of chlorine dioxide dry gas for germ-free avian sterilization practices, augmenting on its use as a fogging agent seen in germ-free mice practices. As described in previous animal laboratory studies, values >1440 ppm-h cycle achieved in this study provided consistently adequate antimicrobial efficacy for the sterilization of germ-free egg facilities.
CONCLUSION: Chlorine dioxide dry gas is a highly effective sterilization solution for germ-free avian egg facilities, providing long-lasting sterility to isolators without the safety concerns associated with other fumigants such as formaldehyde.
METHODS: Polyvinyl chloride (PVC; n = 32) and stainless-steel production isolators (n = 7) were sterilized using approximately 4 h of exposure to chlorine dioxide (PVC isolator) and after approximately 6 h of exposure to chlorine dioxide (stainless-steel production isolator).
RESULTS: Each isolator type was sterilized effectively using chlorine dioxide. Samples collected for microbiological analysis from the isolators after sterilization confirmed that the isolators were sterilized and remained sterile for at least 3 weeks after sterilization.
DISCUSSION: The results of this study highlight the first use of chlorine dioxide dry gas for germ-free avian sterilization practices, augmenting on its use as a fogging agent seen in germ-free mice practices. As described in previous animal laboratory studies, values >1440 ppm-h cycle achieved in this study provided consistently adequate antimicrobial efficacy for the sterilization of germ-free egg facilities.
CONCLUSION: Chlorine dioxide dry gas is a highly effective sterilization solution for germ-free avian egg facilities, providing long-lasting sterility to isolators without the safety concerns associated with other fumigants such as formaldehyde.