MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.diagmicrobio.2018.10.002

Медицинская библиотека · Diagnostic microbiology and infectious disease · 2018

Field evaluation of a dual rapid Human Immunodeficiency Virus and treponemal syphilis rapid test in community-based clinics in Los Angeles and New York.

Stafylis Chrysovalantis, Bristow Claire C, Natoli Lauren J, Salow Katheryn R, Davidson Emma, Granados Yancy, McGrath Mark, Klausner Jeffrey D

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

INTRODUCTION

Dual human immunodeficiency virus/syphilis rapid diagnostic devices can play an important role in prevention efforts. The field performance of the INSTI Multiplex HIV-1/HIV-2/Syphilis Antibody Test (Multiplex) was evaluated.

METHODS

Fingerstick whole blood was tested using the rapid test. A fourth-generation HIV laboratory assay and treponemal-specific laboratory assays were used as reference. Rapid plasma reagin (RPR) was used to stratify treponemal results. Sensitivity and specificity were calculated.

RESULTS

Overall, 274 patients participated. Sensitivity of the Multiplex for detection of HIV was 98.8% (95% CI, 93.4-100%), and specificity was 100% (95% CI, 98.1-100%). Sensitivity for detection of syphilis antibodies was 56.8% (95% CI, 44.7-68.2%), and specificity was 98.5% (95% CI, 95.7-99.7%). Sensitivity for treponemal antibodies improved with increasing RPR and was 100% (95% CI, 78.2-100%) among samples with RPR titers ≥1:8.

CONCLUSIONS

The Multiplex showed excellent performance for detection of HIV antibodies and increasing sensitivity for detection of treponemal antibody with increasing RPR titer.

Для цитирования: Stafylis Chrysovalantis, Bristow Claire C, Natoli Lauren J, Salow Katheryn R, Davidson Emma, Granados Yancy, McGrath Mark, Klausner Jeffrey D. Field evaluation of a dual rapid Human Immunodeficiency Virus and treponemal syphilis rapid test in community-based clinics in Los Angeles and New York.. Diagnostic microbiology and infectious disease. 2018. DOI: 10.1016/j.diagmicrobio.2018.10.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.