DOI 10.1016/j.diagmicrobio.2018.11.018

Медицинская библиотека · Diagnostic microbiology and infectious disease · 2018

Dual rapid test for HIV and syphilis: A laboratory evaluation of the diagnostic accuracy of the Standard Q HIV/Syphilis Combo Test.

Bristow Claire C, Rivera Silver Keith Vargas, Ramos Cordova Lourdes B, Palacios Luz Jazmín Qquellon, Konda Kelika A, Klausner Jeffrey D

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

We evaluated the Standard Q HIV/Syphilis Combo Test (SD Biosensor, South Korea), a dual rapid test using stored sera (N = 400) in a laboratory setting in Lima, Peru. The sensitivity and specificity for HIV antibody detection was 100.0% (95% CI: 98.2-100.0%) and 99.5% (95% CI: 97.2-100.0%), respectively. For treponemal antibody detection the sensitivity and specificity was 97.5% (95%CI:94.3-99.2%) and 100.0% (95%CI:98.2-100.0%), respectively.

Для цитирования: Bristow Claire C, Rivera Silver Keith Vargas, Ramos Cordova Lourdes B, Palacios Luz Jazmín Qquellon, Konda Kelika A, Klausner Jeffrey D. Dual rapid test for HIV and syphilis: A laboratory evaluation of the diagnostic accuracy of the Standard Q HIV/Syphilis Combo Test.. Diagnostic microbiology and infectious disease. 2018. DOI: 10.1016/j.diagmicrobio.2018.11.018.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.