DOI 10.1016/j.jvir.2010.12.016
Comparison of a suspended radiation protection system versus standard lead apron for radiation exposure of a simulated interventionalist.
Аннотация
PURPOSE: To evaluate the radiation protective characteristics of a system designed to enhance operator protection while eliminating weight to the body and allowing freedom of motion.
MATERIALS AND
METHODS
Radiation doses to a mock interventionalist were measured with calibrated dosimeters in a clinical interventional suite. A standard lead apron (SLA; Pb equivalent, 0.5 mm) was compared with a suspended radiation protection system (ZeroGravity; Zgrav) that shields from the top of the head to the calves (except the right arm and left forearm) with a complex overhead motion system that eliminates weight on the operator and allows freedom of motion. Zgrav included a suspended lead apron with increased lead equivalency, greater length, proximal left arm and shoulder coverage, and a wraparound face shield of 0.5 mm Pb equivalency. A 26-cm-thick Lucite stack (ie, mock patient) created scatter during 10 controlled angiography sequences of 120 exposures each. Parameters included a field of view of 40 cm, table height of 94 cm, 124 cm from the tube to image intensifier, 50 cm from the image center to operator, 66 kVp, and 466-470 mA.
RESULTS
Under identical conditions, average doses (SLA vs Zgrav) were 264 versus 3.4 (ratio, 78) to left axilla (P < .001), 456 versus 10.2 (ratio, 45) to left eye (P < .001), 379.4 versus 6.6 (ratio, 57) to right eye (P < .005), and 18.8 versus 1.2 (ratio, 16) to gonad (P < .001).
CONCLUSIONS
Relative to a conventional lead apron, the Zgrav system provided a 16-78-fold decrease in radiation exposure for a mock interventionalist in a simulated clinical setting.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.