Intravascular ultrasound and histology in in vitro assessment of iliac artery angioplasty.
Аннотация
PURPOSE: Intravascular ultrasound (IVUS) was used to assess in vitro the morphologic and quantitative effects of balloon angioplasty (PTA) of the iliac artery.
METHODS
Forty human iliac arteries (>/= 30% area stenosis) were studied with IVUS in vitro before and after PTA and the findings were validated with histology.
RESULTS
The sensitivity of IVUS for dissection was 74% and for media rupture 59%. The incidence of vascular damage was higher when the whole segment was analyzed rather than the target site alone. Dissections occurred at the thinnest region of the plaque, unrelated to plaque calcification. Following PTA, quantitative changes at the target site were greater compared with the overall data derived from all cross-sections. The increase in lumen area was caused solely by an increase in vessel area.
CONCLUSIONS
IVUS is sensitive in detecting dissections, which occurred irrespective of calcification at the thinnest region of the plaque. The increase in lumen area after PTA was caused by stretching of the vessel.