Percutaneous introducibility of the expandable vascular sheath system and injury potential of balloon-assisted thrombectomy: preliminary In vivo results.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Brossmann J, Haghighi P, Bookstein J J

Аннотация

PURPOSE: To test the percutaneous introducibility of the expandable vascular sheath (EVS) system and the safety of percutaneous balloon-assisted thrombectomy.

METHODS: The EVS was inserted directly (n = 9) or through a 9.5 Fr regular vascular introducer sheath (n = 9) into the femoral arteries and veins and carotid arteries in four dogs (18-21 kg). Balloon-assisted thrombectomies were simulated in iliac arteries. Histologic examinations were done at sites of funnel deployment immediately (n = 4) and 25 days (n = 8) after the intervention.

RESULTS: The EVS was successfully introduced into six of nine vessels by a direct percutaneous approach. Balloon-assisted thrombectomy using the EVS device caused localized intimal denudation, disruption of the internal elastic lamina, and mild hemorrhages into the media; one arterial dissection at the site of funnel deployment was seen. All indirect insertions and funnel deployments were successful. Twenty-five days after the experiments, intimal hyperplasia was noted in all cases.

CONCLUSION: Percutaneous balloon-assisted thrombectomy may cause mild vascular injuries. Direct percutaneous introduction of the EVS device cannot be recommended yet.