DOI 10.1016/j.ajog.2008.02.023
Validation of a real-time urodynamic measure of urinary sensation.
Аннотация
OBJECTIVE
The purpose of this study was to test the feasibility and validity of a continuous measurement of urinary sensation during cystometry.
STUDY
DESIGN
Subjects continuously recorded their level of urinary sensation during cystometry with the use of urodynamic diagnoses and responses to the Medical Epidemiologic and Social Aspects of Aging (MESA) and Urinary Distress Inventory (UDI) scales. Trends in urinary sensation recordings by increasing percentage of maximum cystometric capacity (MCC) were captured and compared with the use of growth curves.
RESULTS
Fifty-one women participated: 18 patients had detrusor overactivity incontinence (DOI); 15 patients had urodynamic stress incontinence; 9 patients had mixed incontinence, and 9 patients did not demonstrate incontinence. In the volume range between 35% and 75% of MCC, the mean sensation level was higher in the DOI group than the other groups (P < .04). Urge sensation at 50% of MCC correlated with UDI and MESA urge subscales (rho = 0.34, p < .03 and rho = 0.39, p < .02).
CONCLUSION
This study demonstrates that the Urgeometer measurement is feasible and correlates with urodynamic diagnosis and the severity and bother from urge incontinence.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.