DOI 10.1161/circulationaha.111.053744

Медицинская библиотека · Circulation · 2012

Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee.

Birnie David H, Parkash Ratika, Exner Derek V, Essebag Vidal, Healey Jeffrey S, Verma Atul, Coutu Benoit, Kus Teresa, Mangat Iqwal, Ayala-Paredes Felix, Nery Pablo, Wells George, Krahn Andrew D

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Approximately 268,000 Fidelis leads were implanted worldwide until distribution was suspended because of a high rate of early failure. Careful analyses of predictors of increased lead failure hazard are required to help direct future lead design and also to inform decision making on lead replacement. We sought to perform a comprehensive analysis of all potential predictors in a multicenter study.

METHODS AND

RESULTS

A total of 3169 Sprint Fidelis leads were implanted in 11 centers with a total of 251 failures. Lead failure rates at 3, 4, and 5 years were 5.3%, 10.6%, and 16.8%, respectively. The rate of lead failure continues to accelerate (P<0.001). There were 4 independent predictors of failure: center, sex, access vein, and previous lead failure. Women had a higher hazard of failure (hazard ratio 1.51; 95% confidence interval, 1.14-2.04; P=0.005). Both axillary and subclavian access increased the hazard of failure (P=0.007); hazard ratio for axillary was 1.94, (95% confidence interval, 1.23-3.04) and for subclavian 1.63 (95% confidence interval, 1.08-2.46). Previous lead failure increased the hazard of a subsequent Fidelis failure with a hazard ratio of 3.12 (95% confidence interval, 1.80-5.41; P<0.001).

CONCLUSIONS

The rate of Fidelis failure continues to increase over time, with failures approaching 17% at 5 years. Women, patients with leads inserted via the subclavian or axillary vein, and those with a previous lead fracture were at greatest risk of Fidelis failure. Our data suggest that Fidelis replacement should be strongly considered at the time of generator replacement.

Для цитирования: Birnie David H, Parkash Ratika, Exner Derek V, Essebag Vidal, Healey Jeffrey S, Verma Atul, Coutu Benoit, Kus Teresa, Mangat Iqwal, Ayala-Paredes Felix, Nery Pablo, Wells George, Krahn Andrew D. Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee.. Circulation. 2012. DOI: 10.1161/circulationaha.111.053744.

Предметные рубрики MeSH