MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jjcc.2021.01.018

Медицинская библиотека · Journal of cardiology · 2021

Utility of the minimum-incision transsubclavian approach for transcatheter aortic valve replacement on clinical outcomes in patients with small vessel anatomy.

Domoto Satoru, Jujo Kentaro, Yamaguchi Junichi, Otsuki Hisao, Isomura Shogo, Tanaka Kazuki, Saito Chihiro, Inagaki Yusuke, Yamagata Akiko, Hayakawa Minako, Azuma Takashi, Hagiwara Nobuhisa, Niinami Hiroshi

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The optimal approach for patients undergoing transcatheter aortic valve replacement (TAVR), who are contraindicated for a transfemoral (TF) approach, is still controversial. The present study aimed to evaluate the utility of the TAVR via a subclavian artery with a small diameter, by minimal incision and a double Z suture hemostasis technique using 18 Fr DrySeal Flex sheath, namely minimum-incision transsubclavian TAVR (MITS-TAVR), in patients contraindicated for the TF approach.

METHODS

We included consecutive patients who underwent the MITS-TAVR (MITS group; n = 21) and TF-TAVR (TF group; n = 81) using the CoreValve Evolut R/PRO valves and examined the incidence of in-hospital adverse events and post-discharge mortality between the two groups.

RESULTS

The mean body surface area was significantly smaller in the MITS group (1.33 ± 0.04 vs. 1.43 ± 0.02 m2; p = 0.045). The minimal lumen diameter of the femoral artery was significantly smaller in the MITS group (5.01 vs. 6.43 mm; p < 0.01). The lumen diameter of the left subclavian artery (LSA) in the MITS group was 4.97 ± 0.14 mm. The duration of the TAVR procedure to discharge was not significantly different (9.7 ± 2.0 days vs. 13.2 ± 1.0 days; p = 0.239). We did not experience in-hospital death in both groups, and no significant differences were observed in the incidence of major adverse cardiac and cerebrovascular events between the two groups. The post-discharge survival rate was not significantly different between the groups (at 2-year; MITS group vs. TF group = 91.0% vs. 89.0%; p = 0.725).

CONCLUSIONS

The MITS-TAVR using 18 Fr Dryseal Flex sheath was safe and effective and might be a promising alternative approach even in patients with a small body and small LSA diameter, who are contraindicated to the TF approach.

Для цитирования: Domoto Satoru, Jujo Kentaro, Yamaguchi Junichi, Otsuki Hisao, Isomura Shogo, Tanaka Kazuki, Saito Chihiro, Inagaki Yusuke, Yamagata Akiko, Hayakawa Minako, Azuma Takashi, Hagiwara Nobuhisa, Niinami Hiroshi. Utility of the minimum-incision transsubclavian approach for transcatheter aortic valve replacement on clinical outcomes in patients with small vessel anatomy.. Journal of cardiology. 2021. DOI: 10.1016/j.jjcc.2021.01.018.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.