MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00330-017-5211-3

Медицинская библиотека · European radiology · 2017

Optimisation of sonourethrography: the clamp method.

Berná-Mestre Juan de Dios, Balmaceda Thierry, Martínez Diego, Escudero Jose F, Martínez Gloria, García José A, Canteras Manuel, Berná-Serna Juan D

СтатьяПеревод в очереди

Аннотация

PURPOSE: To describe the clamp method for performing retrograde sonourethrography (RSUG) and contrast-enhanced voiding sonourethrography (CE-VSUG) via the transperineal approach in male adults.

MATERIALS AND

METHODS

Prospective study of 113 males (14-86 years) with urethral strictures confirmed by urethrography who received sonourethrography via the clamp method between 2011 and 2015. The characteristic parameters of the quantitative variables were calculated and a comparative analysis of the qualitative variables was conducted using the McNemar test.

RESULTS

RSUG was performed successfully in all the cases (n = 113) and detected 49 cases with anterior urethral strictures; the strictures in the proximal bulbar cone in five of them (10.2%) were not visualised on retrograde urethrography (RUG) (p < 0.05). CE-VSUG was performed successfully in 97 cases and observed posterior urethral strictures in 82; the bladder neck strictures in 6 of them (7.3%) were not observed on voiding cystourethrography (VCUG) (p < 0.05). Retrograde bladder filling was achieved in approximately 6 min.

CONCLUSION

The clamp method enables RSUG and CE-VSUG to be performed simply, effectively and painlessly by a single operator. It also allows the evaluation of cases with urethromeatal alterations (stricture, hypospadias and meatotomy).

KEY POINTS: • The clamp method enables RSUG to be performed simply and painlessly. • The clamp method requires only one operator and allows assessing urethromeatal alterations. • RSUG shows greater capacity for detecting anterior urethral strictures than RUG. • The clamp method achieves retrograde bladder filling in approximately 6 min. • CE-VSUG shows greater capacity for detecting strictures than VCUG.

Для цитирования: Berná-Mestre Juan de Dios, Balmaceda Thierry, Martínez Diego, Escudero Jose F, Martínez Gloria, García José A, Canteras Manuel, Berná-Serna Juan D. Optimisation of sonourethrography: the clamp method.. European radiology. 2017. DOI: 10.1007/s00330-017-5211-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.