Аннотация

PURPOSE: Some clinical studies have suggested that distal radius plates placed distal to the watershed line have the potential to impinge on the traversing flexor tendons. However, the validity of this theory remains unclear. The purpose of this study was to evaluate the quantitative effect of volar plate position on flexor pollicis longus (FPL) tendon friction by measuring the contact pressure between the FPL tendon and the distal edge of the locking plate.

METHODS

We used 7 fresh cadaveric upper extremities without wrist osteoarthritis or any deformity. External loads of 1.5 and 3.0 kgf were applied to the FPL tendon to simulate the pinch function of the thumb. A distal radius volar plate was applied to these cadaveric specimens in various positions relative to the watershed line. We measured contact pressure between the distal plate edge and the FPL tendon using a thin flexible pressure sensor and compared it among various positions of the volar plate for wrist extension angles of 0°, 30°, and 60° and ulnar deviation angles of 0° and 20°.

RESULTS

Under the 30° or 60° wrist extension condition, contact pressure significantly increased when the distal plate edge was placed distal to the watershed line, compared with when it was placed proximal to or at the watershed line.

CONCLUSIONS

Our quantitative results support the theory that plates placed distal to the watershed line have the potential to impinge on the traversing FPL tendon, even when a radius fracture heals anatomically.

CLINICAL RELEVANCE: This study clarifies a mechanism of FPL tendon irritation after volar plate fixation for distal radius fractures.

Для цитирования: Tanaka Yoshitaka, Aoki Mitsuhiro, Izumi Tomoki, Fujimiya Mineko, Yamashita Toshihiko, Imai Tomohito. Effect of distal radius volar plate position on contact pressure between the flexor pollicis longus tendon and the distal plate edge.. The Journal of hand surgery. 2011. DOI: 10.1016/j.jhsa.2011.08.027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.