Performance of the cobas CT/NG test compared to the Aptima AC2 and Viper CTQ/GCQ assays for detection of Chlamydia trachomatis and Neisseria gonorrhoeae.

Van Der Pol Barbara, Liesenfeld Oliver, Williams James A, Taylor Stephanie N, Lillis Rebecca A, Body Barbara A, Nye Mindy, Eisenhut Carol, Hook Edward W

Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

The next-generation amplification test for Chlamydia trachomatis and Neisseria gonorrhoeae (Roche cobas 4800), a fully automated system, was compared head to head, using female samples, to Gen-Probe Aptima Combo 2 and BD ProbeTec using Viper. Endocervical swabs, female urine, and endocervical samples in liquid-based cytology medium were run on at least two of three platforms. A total of 4,316 samples were evaluated, and 281 chlamydial and 69 gonococcal infections were identified. Estimates of sensitivity and specificity were obtained relative to the patient infection standard (PIS) and using latent class analysis (LCA). Chlamydia sensitivity estimates ranged from 86.9 to 95.6% using PIS and 97.6 to 98% using LCA. Specificity was ≥ 99.6% for all sample types. Sensitivity ranged from 95.6 to 100% using PIS and 96.9 to 100% using LCA for the detection of gonococcal infections. Specificity for gonococcal infections was ≥ 99.8%. cobas 4800 performance was equivalent to the comparator assays (all P values, >0.05), and the fully automated system provides high laboratory efficiency.

Для цитирования: Van Der Pol Barbara, Liesenfeld Oliver, Williams James A, Taylor Stephanie N, Lillis Rebecca A, Body Barbara A, Nye Mindy, Eisenhut Carol, Hook Edward W. Performance of the cobas CT/NG test compared to the Aptima AC2 and Viper CTQ/GCQ assays for detection of Chlamydia trachomatis and Neisseria gonorrhoeae.. Journal of clinical microbiology. 2012. DOI: 10.1128/jcm.06481-11.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.