DOI 10.1016/j.diagmicrobio.2023.116022
Development and validation of a fully automated laboratory-developed test for detection of HSV-1/2 and VZV in clinical samples run on the Panther Fusion® system.
Аннотация
The purpose of this study was to develop a laboratory developed test (LDT) for HSV1/2 and VZV to run on fully automated Hologic Panther Fusion® System. The Panther Fusion System is a fully automated walkaway system, providing end-to-end workflow from sample input to DNA/RNA extraction, amplification, automated analysis, and reporting to a laboratory information system (LIS). The LDT was developed and validated on 230 clinical and 20 reference samples (n = 250) and compared to a commercially available kit. Assessment of the analytical and diagnostic performances of the LDT revealed >98% accuracy, sensitivity, and specificity, which is consistent with or better than many of the commercial or laboratory-developed tests available. The advantage of this LDT is that it is designed to perform a single-run full female health screening in parallel with 4 commercially available Hologic kits for Chlamydia trachomatis/Neisseria gonorrhea (CT/NG), Trichomonas vaginalis (TV), Mycoplasma genitalium (MG), and bacterial vaginosis (BV).
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.