Медицинская библиотека · Journal of vascular and interventional radiology : JVIR · 1998

Tissue response to covered Wallstents.

Dolmatch B L, Dong Y H, Trerotola S O, Hunter D W, Brennecke L H, LaBounty R

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To evaluate and compare the healing response related to two types of graft-covered Wallstents (WSs) and an uncovered WS in the canine iliac artery.

MATERIALS AND

METHODS

Eight bare mesh WSs, 10 polyethylene terephthalate interbraided WSs (PET-WSs), and six polytetrafluoroethylene covered WSs (ePTFE-WSs) were placed in the iliac arteries of 12 dogs. Arteriograms were obtained before and after implantation and at explantation. Devices were explanted at 1 month (n = 8), 3 months (n = 6), and 6 months (n = 10) and sent for histologic study.

RESULTS

One ePTFE-covered stent-graft was found to be thrombosed at 3 months; the remaining 23 of 24 implants remained patent to the time of explantation. The WS and ePTFE-WS both generated a uniform myointimal cell response without inflammation. The PET-WS induced a fibrous luminal response with substantial foreign body-type inflammatory reaction around the PET fibers. Although neointima associated with the PET-WS appeared thicker than that associated with either the ePTFE-WS or the bare WS, none of the patent implants developed greater than 50% angiographic narrowing.

CONCLUSION

The PET-WS induced chronic inflammation, a response not seen with either the WS or ePTFE-WS. This may explain the occurrence of pain and/or fever in human studies of arterial PET endoluminal stent-grafts. Patency for all three implants was excellent in this model.

Для цитирования: Dolmatch B L, Dong Y H, Trerotola S O, Hunter D W, Brennecke L H, LaBounty R. Tissue response to covered Wallstents.. Journal of vascular and interventional radiology : JVIR. 1998. DOI: 10.1016/s1051-0443(98)70302-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.