PivNG primers and probes set used in the cobas omni Utility Channel is a reliable supplemental test for detection of Neisseria gonorrhoeae in oropharyngeal, urogenital and rectal specimens collected in cobas PCR Media.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Hopkins Mark, Arcenas Rodney, Couto-Parada Xose, Lewinski Michael, Njoya Merlin, Perinpanathan Deva, Sheriff Rizwaan, Hansra Avneet, Singh Selena
Аннотация
OBJECTIVE: To evaluate the clinical performance of the novel PivNG primers and probes set (PivNG test) used in the cobas omni Utility Channel for supplemental testing of Neisseria gonorrhoeae (NG).
METHODS: Oropharyngeal, urogenital and rectal samples were self-collected during routine testing at Barts Health sexual health clinics, London, UK. Samples were tested by the cobas CT/NG test and PivNG cobas omni Utility Channel test on cobas 6800/8800 Systems. Supplemental testing was carried out with the Xpert CT/NG test. PivNG overall percent agreements, positive percent agreements (PPAs)/negative percent agreements (NPAs) and positive/negative predictive values were calculated for each sample type. Microscopy and/or culture data were included for a randomised subset of concordant/discordant results, and a composite reference standard (cobas CT/NG, Xpert CT/NG and culture results) adjusted for partial verification bias was used to determine PivNG PPA and NPA.
RESULTS: Of 447 evaluable samples with valid results from all three assays (cobas CT/NG, PivNG and Xpert CT/NG), 288 (64.4%) were NG-positive by both PivNG and cobas CT/NG; 117 (26.2%) were NG-negative in both tests; and 42 (9.4%) had discordant results (with NG-negative supplementary Xpert) CT/NG results in 40/42 instances). Of 19 PivNG/Xpert CT/NG-discordant samples, 11 were confirmed NG-positive by microscopy and/or culture results. PivNG PPA and NPA were 100% and 91% for oropharyngeal swabs, 100% and 100% for vaginal swabs, 100% and 100% for male urine samples, and 100% and 97% for rectal swabs, respectively, compared with the partially adjusted composite reference standard.
CONCLUSIONS: PivNG is a reliable supplementary test with high sensitivity for confirming NG infection when used in conjunction with the cobas CT/NG test and samples collected in cobas PCR Media. Moreover, the PivNG test offers a convenient, high-throughput solution for supplemental NG testing of various sample types, with the potential to reduce the number of indeterminate reports.
METHODS: Oropharyngeal, urogenital and rectal samples were self-collected during routine testing at Barts Health sexual health clinics, London, UK. Samples were tested by the cobas CT/NG test and PivNG cobas omni Utility Channel test on cobas 6800/8800 Systems. Supplemental testing was carried out with the Xpert CT/NG test. PivNG overall percent agreements, positive percent agreements (PPAs)/negative percent agreements (NPAs) and positive/negative predictive values were calculated for each sample type. Microscopy and/or culture data were included for a randomised subset of concordant/discordant results, and a composite reference standard (cobas CT/NG, Xpert CT/NG and culture results) adjusted for partial verification bias was used to determine PivNG PPA and NPA.
RESULTS: Of 447 evaluable samples with valid results from all three assays (cobas CT/NG, PivNG and Xpert CT/NG), 288 (64.4%) were NG-positive by both PivNG and cobas CT/NG; 117 (26.2%) were NG-negative in both tests; and 42 (9.4%) had discordant results (with NG-negative supplementary Xpert) CT/NG results in 40/42 instances). Of 19 PivNG/Xpert CT/NG-discordant samples, 11 were confirmed NG-positive by microscopy and/or culture results. PivNG PPA and NPA were 100% and 91% for oropharyngeal swabs, 100% and 100% for vaginal swabs, 100% and 100% for male urine samples, and 100% and 97% for rectal swabs, respectively, compared with the partially adjusted composite reference standard.
CONCLUSIONS: PivNG is a reliable supplementary test with high sensitivity for confirming NG infection when used in conjunction with the cobas CT/NG test and samples collected in cobas PCR Media. Moreover, the PivNG test offers a convenient, high-throughput solution for supplemental NG testing of various sample types, with the potential to reduce the number of indeterminate reports.