Diagnosis of preeclampsia and fetal growth restriction with the sFlt-1/PlGF ratio: Diagnostic accuracy of the automated immunoassay Kryptor®.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Dröge Lisa Antonia, Höller Alice, Ehrlich Laura, Verlohren Stefan, Henrich Wolfgang, Perschel Frank Holger
Аннотация
OBJECTIVE: We aimed to characterize the diagnostic accuracy of the Kryptor® assay for sFlt-1 and PlGF in maternal serum samples of uneventful singleton pregnancies and subjects with preeclampsia (PE) and PE-related outcomes such as fetal growth restriction (FGR). Longitudinal reference ranges of the sFlt-1 and PlGF level in the course of normal pregnancies were generated.
METHODS: A cohort of subjects with PE and PE-related outcomes including FGR in the third trimester was compared to a cohort of women with uneventful outcome. Serum levels of sFlt-1, PlGF level as well as the sFlt-1/PlGF ratio was analysed with the Kryptor® assay and compared between the case- and control groups. Cut-off values were generated and diagnostic accuracy examined.
RESULTS: Longitudinal reference ranges of the sFlt-1 and PlGF level in healthy pregnancies were in line with those levels measured with other immunoassays. Comparison of the sFlt-1/PlGF ratio between PE-related outcomes including FGR or PE and healthy controls showed a high diagnostic accuracy with an area under the curve (AUC) of 0.917 for PE-related outcomes and 0.919 for PE.
METHODS: A cohort of subjects with PE and PE-related outcomes including FGR in the third trimester was compared to a cohort of women with uneventful outcome. Serum levels of sFlt-1, PlGF level as well as the sFlt-1/PlGF ratio was analysed with the Kryptor® assay and compared between the case- and control groups. Cut-off values were generated and diagnostic accuracy examined.
RESULTS: Longitudinal reference ranges of the sFlt-1 and PlGF level in healthy pregnancies were in line with those levels measured with other immunoassays. Comparison of the sFlt-1/PlGF ratio between PE-related outcomes including FGR or PE and healthy controls showed a high diagnostic accuracy with an area under the curve (AUC) of 0.917 for PE-related outcomes and 0.919 for PE.
Предметные рубрики MeSH
AdultArea Under CurveAutomation, LaboratoryBiomarkersCase-Control StudiesFemaleFetal Growth RetardationHumansImmunoassayPlacenta Growth FactorPre-EclampsiaPredictive Value of TestsPregnancyPregnancy Trimester, ThirdROC CurveReference StandardsReproducibility of ResultsVascular Endothelial Growth Factor Receptor-1Young Adult