Validation of the Omron MIT Elite blood pressure device in a pregnant population with large arm circumference.
Аннотация
OBJECTIVE
The aim of this study was to evaluate the accuracy of the Omron MIT Elite automated device in pregnant women with an arm circumference of or above 32 cm, using the British Hypertension Society validation protocol.
METHODS
Blood pressure was measured sequentially in 46 women of any gestation requiring the use of a large cuff (arm circumference ≥32 cm) alternating between the mercury sphygmomanometer and the Omron MIT Elite device.
RESULTS
The Omron MIT Elite achieved an overall D/D grade with a mean of the device-observer difference being 7.17±6.67 and 9.31±6.59 for systolic and diastolic blood pressure respectively. Interobserver accuracy was 94.6% for systolic and 95% for diastolic readings within 5 mmHg.
CONCLUSION
The Omron MIT Elite overestimates blood pressure and has failed the British Hypertension Society protocol requirements. Therefore, it cannot be recommended for use in pregnant women with an arm circumference of or above 32 cm.