MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3855/jidc.15279

Медицинская библиотека · Journal of infection in developing countries · 2022

Comparison of two commercial multiplex PCR assays for the detection of sexually transmitted infections.

Yassin Ruba, Hanna Jessica, El Bikai Rana, Mohtar Jad, El Chaar Mira

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Multiplex molecular panels are replacing conventional methods for the detection of sexually transmitted infections. In the current study, we evaluated the performance of two commercial multiplex assays, EUROArray STI and Allplex STI essential assays, for detecting six sexually transmitted infections.

METHODOLOGY: The diagnostic performance of the EUROArray STI and Allplex STI essential assays was evaluated against a panel of 105 positive DNA samples identified by in-house real-time PCR assays including Ureaplasma parvum, Ureaplasma urealyticum, Mycoplasma hominis, Trichomonas vaginalis, Chlamydia trachomatis, and Neisseria gonorrhea. Samples from healthy subjects, negative for any microorganism, were used as negative controls.

RESULTS

Of the 105 positive specimens, 103 (98%) were tested positive by Allplex and 102 (97%) by EUROArray. Among the 51 negative samples that were tested by in house assay, 48 (94%) were tested negative by Allplex assay and 43 (84%) by EUROArray assay. The overall sensitivity of EUROArray and Allplex were 97.1% and 98.1% with an accuracy of 92.9% and 96.7%, respectively. The overall assay specificity was 94.1% for Allplex assay and 84.3% for EUROArray assay, The sensitivity of both kits to all targeted microorganisms ranged from 55.6% to 100%, with the lowest sensitivity noted for Trichomonas vaginalis.

CONCLUSIONS

Diagnostic performance varies depending on the method used to detect the targeted pathogens, the assay manipulation, and the cost. This study showed sensitivity, specificity, and accuracy characteristics for two kits commonly used to detect STIs, which will guide the choice for an appropriate multiplex PCR platform.

Для цитирования: Yassin Ruba, Hanna Jessica, El Bikai Rana, Mohtar Jad, El Chaar Mira. Comparison of two commercial multiplex PCR assays for the detection of sexually transmitted infections.. Journal of infection in developing countries. 2022. DOI: 10.3855/jidc.15279.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.