DOI 10.1053/j.ajkd.2005.08.001
Efficiency of the Genius batch hemodialysis system with low serum solute concentrations: the case of lithium intoxication therapy.
Аннотация
BACKGROUND
The Genius batch system consists of a 90-L closed reservoir, from which fresh dialysate is extracted at the top and to which spent dialysate is returned at the bottom. It was shown in long-term hemodialysis patients that almost the entire amount of unspent dialysate can be used before contamination of fresh with spent dialysate occurs. Separation is caused by differences in density, partly because of the presence of uremic solutes in spent dialysate. The question is raised whether this separation can be maintained during dialysis of patients who experience an intoxication without renal failure.
METHODS
A patient intoxicated with lithium was dialyzed using the Genius system, prepared at 37 degrees C, during 300 minutes. With dialysate flow set at 300 mL/min (5 mL/s) and in the absence of mixing, urea is not expected at the inlet dialysate tubing before minute 300.
RESULTS
In the dialysate inlet tubing, an abrupt increase in lithium and urea concentrations was observed 210 minutes after the start of the session, reflecting contamination of fresh with spent dialysate. At minute 210, only 60.9 L of 90 L of dialysate had crossed the dialyzer. In a control dialysis treatment in a patient with marked renal failure, this mixing occurred only at 300 minutes.
CONCLUSION
In the present observation, it is shown that during Genius dialysis in a patient without renal failure, an earlier contamination of fresh with spent dialysate can occur, compared to conditions of renal failure.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.