Non-invasive cardiac output measurement by electrical cardiometry and M-mode echocardiography in the neonate: a prospective observational study of 136 neonatal infants.
Аннотация
BACKGROUND
Electrical cardiometry (EC) is a continuous, non-invasive method for measuring cardiac output (CO). This study investigates the correlation and consistency of CO values in newborns obtained by using EC and M-mode echocardiography (Teichholz formula).
METHODS
In this prospective observational study, simultaneous measurement of CO was implemented with EC (COec) and M-mode echocardiography (COm) in neonates. The absolute values of CO measured by the two methods were converted to Z-scores. Following that, Pears's correlation analyses and the Bland-Altman index were employed to analyze the correlation and consistency of COec Z-scores and COm Z-scores.
RESULTS
A total of 136 neonates (93 preterm infants) were enrolled in this study, and EC and M-mode echocardiography comparative studies were conducted 155 times. The mean value of COec and COm demonstrated significant statistical differences (P<0.001). A moderate correlation (r=0.601; P<0.001) was found between the two methods. The Bland-Altman index value was 3.2%, which remained less than 5% in the low birth weight (LBW) (2.1%), non-LBW (3.4%), spontaneous respiration (3.1%), nasal continuous positive airway pressure (nCPAP) (4.0%), mechanical ventilation (2.9%), hemodynamic significance of the patent ductus arteriosus (hsPDA) (4.3%), and non-hsPDA (3.7%) groups, respectively.
CONCLUSIONS
Although the absolute values of CO measured by EC and M-mode echocardiography were not interchangeable, the distribution of CO in EC and M-mode echocardiography was similar.