Homemade Ultrasound Stand-Off Gel Pads: A Multi-Parameter Performance Study.

William Muhammad Fiqqrie De Gracious Bin Mohd Firdaus, Jamil Nur Arisha Binti, Leong Sook Sam, Supar Rafidah, Ab Mumin Nazimah, Ishak Ahmad Razali, Idzam Nurul Syafiqah Binti Noor, Ajis Mohd Nazzihan Md, Vijayananthan Anushya

СтатьяПеревод в очереди

Аннотация

AIM: Stand-off gel pads made of natural ingredients may provide a low-cost option in clinical ultrasound imaging. This study investigated the performance and safety of homemade pads.

METHODS

Thirty stand-off gel pads were made of aloe vera, xanthan gum and cucumber. Testing of each was performed to assess mechanical failure, moisture loss, contamination rate and image quality. Mechanical strength was measured using a tensile testing machine; drying rate was calculated using the moisture loss formula; contamination was evaluated through colony counting; and image quality was assessed using a chicken breast phantom.

RESULTS

There was a significant difference in compressive force resistance between the stand-off gel pads (p = 0.044). The force required was higher for those made of xanthan gum compared with cucumber (p = 0.047). There was a significant difference in moisture loss (p < 0.001). Aloe vera gel exhibited a higher moisture loss (9.65% ± 0.15%) compared with xanthan gum (5.94% ± 0.11%) and cucumber (5.81% ± 0.11%) (p < 0.001). There was a significant difference in axial resolution (p = 0.037), lateral resolution (p = 0.029) and striations in the chicken breast parenchyma (p = 0.037). Speckle artefacts are presented in xanthan-gum-based stand-off gel pad. There was a borderline significant difference of artefact presence in images produced using all pads (p = 0.050). Microbial counts were too low to be quantified.

CONCLUSION

A cucumber stand-off gel pad is a practical alternative for clinical use owing to its structural integrity, retained moisture and low microbial contamination. The image quality produced using the pad was comparable with the conventional glove-filled-with-water stand-off method.

Для цитирования: William Muhammad Fiqqrie De Gracious Bin Mohd Firdaus, Jamil Nur Arisha Binti, Leong Sook Sam, Supar Rafidah, Ab Mumin Nazimah, Ishak Ahmad Razali, Idzam Nurul Syafiqah Binti Noor, Ajis Mohd Nazzihan Md, Vijayananthan Anushya. Homemade Ultrasound Stand-Off Gel Pads: A Multi-Parameter Performance Study.. Australasian journal of ultrasound in medicine. 2026. DOI: 10.1002/ajum.70044.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.