DOI 10.1007/s10096-020-04135-9

Медицинская библиотека · European journal of clinical microbiology & infectious diseases : official publication of the European Society of Clinical Microbiology · 2021

Comparison between Abbott m2000 RealTime and Alinity m STI systems for detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Mycoplasma genitalium.

Herrmann Björn, Malm Karin

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

The new Abbott Alinity m STI Assay was compared with Abbott m2000 RealTime PCR. For Chlamydia trachomatis, 26 (7.5%) of 347 samples were positive in the Alinity assay and 24 (6.9%) in the m2000 assay. Corresponding figures for Neisseria gonorrhoeae were 23 (6.6%) and 17 (4.9%). For Mycoplasma genitalium, 22 (7.9%) of 279 samples were positive in the Alinity assay and 18 (6.5%) in the m2000 assay, for which DNA extraction was performed on an m2000sp instrument combined with in-house real-time PCR. The Alinity assay has at least the same sensitivity as the m2000 assay. The specificity was evaluated by discrepancy analysis.

Для цитирования: Herrmann Björn, Malm Karin. Comparison between Abbott m2000 RealTime and Alinity m STI systems for detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Mycoplasma genitalium.. European journal of clinical microbiology & infectious diseases : official publication of the European Society of Clinical Microbiology. 2021. DOI: 10.1007/s10096-020-04135-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.