MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1583/1074-6218(1997)004<0124:tyewtw>2.0.co;2

Медицинская библиотека · Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery · 1997

Three-year experience with the White-Yu Endovascular GAD Graft for transluminal repair of aortic and iliac aneurysms.

White G H, Yu W, May J, Waugh R, Chaufour X, Harris J P, Stephen M S

СтатьяПеревод в очереди

Аннотация

PURPOSE: To report a > 3-year experience with a modular, balloon-expandable endovascular graft used for aneurysm exclusion in the aorta and other arteries.

METHODS

The customized White-Yu Endovascular GAD Graft, a woven polyester prosthesis with an intrinsic Elgiloy wire graft attachment system along the body of the graft, is a flexible endograft design available in straight, tapered, and bifurcated versions that can be delivered transluminally through 18F to 24F sheaths.

RESULTS

Since July 1993, 93 patients have received the White-Yu endograft for treatment of 76 abdominal aortic, 3 thoracic aortic, 13 iliac, and 1 popliteal aneurysms. Of the 79 aortic procedures, 39 involved straight tube grafts, 20 were tapered aortoiliac models, and 20 were bifurcated devices. Success rates for tube grafts were 81% in the abdominal aorta and 100% for the thoracic aorta; 5 primary endoleaks (14%) and 2 conversions to surgery (5.6%) occurred with this graft type. Aortoiliac grafts were deployed successfully in 95% (19/20) of cases with 1 conversion (5%) due to thrombosis. Seventy-five percent of the bifurcated endograft procedures were successful, with 4 conversions (20%) for technical failures and 1 graft thrombosis. Four additional endografts were deployed to treat two primary and two secondary endoleaks in tube graft patients. Two access-related arterial injuries were treated surgically. There was one case of embolus to the distal femoral artery but no microembolization. Overall perioperative (30-day) mortality was 3.1%. Over a mean 18-month follow-up (range 2 to 39), no late graft thrombosis, stenosis, or graft migration has been seen on CT scans or X ray. Endoleak has not been detected in any aortoiliac or bifurcated graft. Aneurysm size has diminished consistently in successfully treated cases.

CONCLUSIONS

The White-Yu endograft appears to offer a safe, efficacious, and minimally invasive means of excluding aneurysms from the circulation. Improvements in patient selection, surgical techniques, and equipment have reduced the incidence of endoleak and conversion to open repair over the course of the evaluation.

Для цитирования: White G H, Yu W, May J, Waugh R, Chaufour X, Harris J P, Stephen M S. Three-year experience with the White-Yu Endovascular GAD Graft for transluminal repair of aortic and iliac aneurysms.. Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery. 1997. DOI: 10.1583/1074-6218(1997)004<0124:tyewtw>2.0.co;2.