Effect of Phototherapy on the Reliability of Transcutaneous Bilirubin Devices in Term and Near-Term Infants: A Systematic Review and Meta-Analysis.
Аннотация
BACKGROUND
Transcutaneous bilirubin (TcB) devices are commonly used for screening of hyperbilirubinemia in term and near-term infants not exposed to phototherapy. However, the accuracy of TcB devices in infants exposed to phototherapy is unclear.
OBJECTIVES
To conduct a systematic review of studies comparing TcB devices with total serum bilirubin (TSB) in infants receiving phototherapy or in the postphototherapy phase.
METHODS
MEDLINE, EMBASE, Cochrane Library, CINAHL and Scopus databases (from inception to May 8, 2014) were searched. Additional citations were identified from the bibliography of selected articles and from the abstracts of conference proceedings. The studies were included if they compared TcB results with TSB in term and near-term infants during phototherapy or after discontinuation of phototherapy. Two reviewers independently assessed studies for inclusion, and discrepancies were resolved with consensus. Risk of bias was assessed using the QUADAS-2 tool.
RESULTS
Fourteen studies were identified. The pooled estimates of correlation coefficients (r) during phototherapy were: covered sites 0.71 (95% CI 0.64-0.77, 11 studies), uncovered sites 0.65 (95% CI 0.55-0.74), 8 studies), forehead 0.70 (95% CI 0.64-0.75, 12 studies) and sternum 0.64 (95% CI 0.43-0.77, 5 studies). Two studies also provided results as Bland-Altman difference plots (mean TcB-TSB differences -29.2 and 30 µmol/l, respectively). The correlation coefficient improved marginally in the postphototherapy phase (r = 0.72, 95% CI 0.64-0.78, 4 studies).
CONCLUSION
We found a moderate correlation between TcB and TSB during phototherapy with a marginal improvement in the postphototherapy phase. Further research is needed before the use of TcB devices can be recommended for these settings.