Quality assessment of confocal microscopy slide-based systems: instability.

Zucker Robert M

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

All slide-based fluorescence cytometry detections systems basically include an excitation light source, intermediate optics, and a detection device (CCD or PMT). Occasionally, this equipment becomes unstable, generating unreliable and inferior data.

METHODS

A number of tests have been devised to evaluate equipment performance and instability. The following four instability tests are described: galvanometer scanning, stage drift, correct wavelength spectral detection, and long-term laser power.

RESULTS

Quality assurance tests revealed that a confocal microscope can become unstable in the following parameters, yielding inaccurate data: laser power, PMTs functionality, spectrophotometer accuracy, galvanometer scanning and laser stability, and stage drift. Long-term laser power stability has been observed to vary greatly.

CONCLUSIONS

Confocal systems can become unstable in the following parameters: long-term laser power, galvanometer scanning, spectrophotometer accuracy, and stage stability. Instability in any of these parameters will affect image quality. Laser power fluctuations result from either a defective Acousto-optic tunable filter or improper heat dissipation. Spectrophotometer instability will generate unreliable spectra data, extra light reflections, and poor image quality. Galvanometer scanning instability yields poor image quality while microscope stage drift results in a sample going out of the plane of focus. With minor modifications, these tests may be applicable to other slide-based systems.

Для цитирования: Zucker Robert M. Quality assessment of confocal microscopy slide-based systems: instability.. Cytometry. Part A : the journal of the International Society for Analytical Cytology. 2006. DOI: 10.1002/cyto.a.20313.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.