Percutaneous closure of ventricular septal defects in children: key parameters affecting patient radiation exposure.
Аннотация
BACKGROUND
Ventricular septal defect (VSD) transcatheter closure is gaining worldwide popularity despite its complexity. Reports on key factors affecting radiation exposure in children are scarce.
AIMS: This clinical study is the first to comprehensively analyze the impact of all relevant parameters on children's radiation exposure during VSD interventional closures.
METHODS
Between March 2016 and August 2019, all pediatric VSD cases percutaneously treated at a reference center for interventional congenital cardiology and equipped with a single-plane Innova 2100 X-ray unit were retrospectively reviewed. Multiple linear regression was performed to investigate the impact of clinical, technical, and procedural parameters on patients' radiation exposure assessed using total air kerma area product (PKA,T).
RESULTS
A total of 85 patients were included in this study and 82.4% had perimembranous defects. Device implantation was successful in 96.5% of cases. The procedure lasted for a median of 60 min with a median PKA,T of 19.6 Gy.cm2 (range, 1.1 to 244.8 Gy.cm2). Patients' weight (B = 1.679, P = 0.01), number of operators (B = 1.561, P = 0.02), device positioning complexity (B = 2.381, P = 0.002), and procedural incidents (B = 2.096, P = 0.008) significantly increased PKA,T. Patients' age (B = 1.053, P = 0.784), device design (B = -1.216, P = 0.780) and approach of delivery (B = -1.119, P = 0.511) did not significantly affect PKA,T.
CONCLUSIONS
Radiation exposure in children undergoing VSD percutaneous closure was highly variable. A higher patient's weight, numbers of operators, complexity in device positioning, and procedural incidents, were identified as key factors increasing patient dose for this kind of intervention.