Evaluation of a dilution method for non-evaluable results in the detection of Chlamydia trachomatis and Neisseria gonorrhoeae with the Cobas 4800 platform.
СтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Parra-Sánchez Manuel, García-Rey Silvia, Zakariya-Yousef Breval Ismail, Sierra-Atienza Celestina, Bernal-Martínez Samuel, Palomares-Folía José Carlos
Аннотация
INTRODUCTION: A variable percentage of samples analysed using the Cobas 4800 assay can give an invalid result by PCR inhibition or erroneous due to incorrect DNA extraction with the Cobas 4800 CT/NG test.
METHOD: An analysis was performed using the vortex agitation and dilution protocol on the original sample (swab or urine) for a total of 116 samples. In order to analyse the sensitivity of this method, 100 samples (swabs and urine) with known results were retested.
RESULTS: A total of 98.3% (114/116) of the samples analysed were resolved with this protocol with 100% agreement after reviewing clinical data, Gram stain, and other samples analysed in parallel from the same patient.
DISCUSSION: The data indicate no loss of sensitivity with this protocol; thus Cobas 4800 users could use this method without the need for alternative methods.
METHOD: An analysis was performed using the vortex agitation and dilution protocol on the original sample (swab or urine) for a total of 116 samples. In order to analyse the sensitivity of this method, 100 samples (swabs and urine) with known results were retested.
RESULTS: A total of 98.3% (114/116) of the samples analysed were resolved with this protocol with 100% agreement after reviewing clinical data, Gram stain, and other samples analysed in parallel from the same patient.
DISCUSSION: The data indicate no loss of sensitivity with this protocol; thus Cobas 4800 users could use this method without the need for alternative methods.