DOI 10.1177/1071100713496223

Медицинская библиотека · Foot & ankle international · 2013

Biomechanical comparison of 4 different lateral plate constructs for distal fibula fractures.

Eckel Tobin T, Glisson Richard R, Anand Prashanth, Parekh Selene G

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Displaced lateral malleolar fractures are often treated with reduction and surgical stabilization. However, there has not been a comprehensive laboratory comparison to determine the most appropriate device for treating these patients. This study subjected a range of contemporary lateral fibular plates to a series of mechanical tests designed to reveal performance differences.

METHODS

Forty fresh frozen lower extremities were divided into 4 groups. A Weber B distal fibula fracture was simulated with an osteotomy and stabilized using 1 of 4 plate systems: a standard Synthes one-third tubular plate with interfragmentary lag screw, a Synthes LCP locking plate with lag screw, an Orthohelix MaxLock Extreme low-profile locking plate with lag screw, or a TriMed Sidewinder nonlocking plate. Controlled monotonic bending and cyclic torsional loading were applied and bending stiffness, torsional stiffness, and fracture site motion were quantified. Resistance to cyclic torsional loading was determined by quantifying the number of loads withstood before excessive rotation occurred. Correlation between bone mineral density and each of the mechanical measures was determined.

RESULTS

There was no difference in angulation or bending stiffness between plates. All plates except the LCP showed greater lateral deflection than in the other bending directions. Bending stiffness was lowest in lateral distal fragment deflection for all 4 plates. There was a positive correlation between bone mineral density and bending stiffness for all plate types. There was no difference in fracture site rotation between plate types in internal or external torsion, but internal rotation of the distal fragment consistently exceeded external rotation. Torsional stiffness in external rotation exceeded stiffness in internal rotation in nearly all specimens. LCP plates performed relatively poorly under cyclic torsion.

CONCLUSIONS

Significant differences in plate performance were not demonstrated. The effects of bone quality variability and differences in interfragmentary screw purchase resulted in data dispersion that confounded absolute ranking of plate performance.

CLINICAL RELEVANCE: Identification of an optimal lateral fibular plating system has the potential to improve the clinical outcome of malleolar fracture fixation, particularly when patient conditions are unfavorable.

Для цитирования: Eckel Tobin T, Glisson Richard R, Anand Prashanth, Parekh Selene G. Biomechanical comparison of 4 different lateral plate constructs for distal fibula fractures.. Foot & ankle international. 2013. DOI: 10.1177/1071100713496223.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.