The esophageal detector device: accuracy and reliability in difficult airway settings.

Kapsner C E, Seaberg D C, Stengel C, Ilkhanipour K, Menegazzi J

Клиническое исследованиеСтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

The esophageal detector device (EDD) recently has been found to assess endotracheal (ET) tube placement accurately. This study describes the reliability of the EDD in determining the position of the ET tube in clinical airway situations that are difficult.

METHODS

This was a prospective, randomized, single-blinded, controlled laboratory investigation. Two airway managers (an emergency-medicine attending physician and a resident) determined ET-tube placement using the EDD in five swine in respiratory arrest. The ET tube was place in the following clinical airway situations: 1) esophagus; 2) esophagus with 1 liter of air instilled; 3) trachea; 4) trachea with 5 ml/kg water instilled; and 5) right mainstem bronchus. Anatomic location of the tube was verified by thoracotomy of the left side of the chest.

RESULTS

There was 100% correlation between the resident and attending physician's use of the EDD. The EDD was 100% accurate in the determining tube placement in the esophagus, in the esophagus with 1 liter of air instilled, in the trachea, and in the right mainstem bronchus. The airway managers were only 80% accurate in detecting tracheal intubations when fluid was present.

CONCLUSIONS

The EDD is an accurate and reliable device for detecting ET-tube placement in most clinical situations. Tube placement in fluid-filled trachea, lungs, or both, which occurs in pulmonary edema and drowning, may not be detected using this device.

Для цитирования: Kapsner C E, Seaberg D C, Stengel C, Ilkhanipour K, Menegazzi J. The esophageal detector device: accuracy and reliability in difficult airway settings.. Prehospital and disaster medicine. 1996. DOI: 10.1017/s1049023x00042369.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.