Could electrohysterography be the solution for external uterine monitoring in obese women?
Аннотация
OBJECTIVE
To evaluate the influence of maternal obesity on the performance of external tocodynamometry and electrohysterography.
STUDY
DESIGN
In a 2-hour measurement during term labor, uterine contractions were simultaneously measured by electrohysterography, external tocodynamometry, and intra-uterine pressure catheter. The sensitivity was compared between groups based on obesity (non-obese/obese/morbidly obese) or uterine palpation (good/moderate/poor), and was correlated to maternal BMI and abdominal circumference.
RESULT: We included 14 morbidly obese, 18 obese, and 20 non-obese women. In morbidly obese women, the median sensitivity was 87.2% (IQR 74-93) by electrohysterography and 45.0% (IQR 36-66) by external tocodynamometry (p < 0.001). The sensitivity of electrohysterography appeared to be non-influenced by obesity category (p = 0.279) and uterine palpation (p = 0.451), while the sensitivity of tocodynamometry decreased significantly (p = 0.005 and p < 0.001, respectively). Furthermore, the sensitivity of both external methods was negatively correlated with obesity parameters, being non-significant for electrohysterography (range p-values 0.057-0.088) and significant for external tocodynamometry (all p-values < 0.001).
CONCLUSIONS
Electrohysterography performs significantly better than external tocodynamometry in case of maternal obesity.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.