Obstructive Shock Because of the Incorrect Setup of a Digital Drainage System After Pulmonary Lobectomy.

Thind Dilraj, Jamshidi Shirin, Vitesnikova Jana

Клинический случайСтатьяПеревод в очереди

Аннотация

We report a case of life-threatening obstructive shock from extreme negative intrapleural pressure that occurred because of the incorrect setup of a digital drainage system after video-assisted thoracic surgery. These devices are increasingly prevalent because of their precision and portability. Those using such devices should be cognizant of their potential pitfalls.

Для цитирования: Thind Dilraj, Jamshidi Shirin, Vitesnikova Jana. Obstructive Shock Because of the Incorrect Setup of a Digital Drainage System After Pulmonary Lobectomy.. The Annals of thoracic surgery. 2021. DOI: 10.1016/j.athoracsur.2021.10.069.