Comparison of the BD Viper System with XTR Technology to the Gen-Probe APTIMA COMBO 2 Assay using the TIGRIS DTS system for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae in urine specimens.

Mushanski Linda M, Brandt Ken, Coffin Nicolette, Levett Paul N, Horsman Gregory B, Rank Elliot L

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Performances of the BD ProbeTec Chlamydia trachomatis (CT)/Neisseria Gonorrhoeae (GC) Q(x) Amplified DNA Assay reagents on a BD Viper System with XTR Technology and APTIMA COMBO 2 Assay reagents on a TIGRIS DTS platform, for detection of both CT and GC were compared.

METHODS

A total of 1018 first-void urine specimens were tested for the presence of CT and GC DNA using the 2 assays.

RESULTS

CT was detected in 143 specimens (14%). Eight specimens exhibited discordant results, and they were divided equally between the 2 assays. Based on the original results, the overall agreement for CT was 99.2%, with 97.1% and 99.5% in agreement with positive and negative specimens, respectively. Cohen's Kappa was 0.967. GC was detected in 27 specimens (2.6%). Two specimens exhibited discordant results, and they were divided equally between the 2 assays. Based on the original results, the overall agreement was 99.8%, with 96.2% and 99.9% in agreement for positive and negative specimens, respectively. Cohen's Kappa was 0.961.

CONCLUSIONS

There was a high level of agreement between the systems for both CT and GC detection.

Для цитирования: Mushanski Linda M, Brandt Ken, Coffin Nicolette, Levett Paul N, Horsman Gregory B, Rank Elliot L. Comparison of the BD Viper System with XTR Technology to the Gen-Probe APTIMA COMBO 2 Assay using the TIGRIS DTS system for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae in urine specimens.. Sexually transmitted diseases. 2012. DOI: 10.1097/olq.0b013e31824f2f5b.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.