Comparison of the BD Viper System with XTR Technology to the Gen-Probe APTIMA COMBO 2 Assay using the TIGRIS DTS system for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae in urine specimens.
Аннотация
BACKGROUND
Performances of the BD ProbeTec Chlamydia trachomatis (CT)/Neisseria Gonorrhoeae (GC) Q(x) Amplified DNA Assay reagents on a BD Viper System with XTR Technology and APTIMA COMBO 2 Assay reagents on a TIGRIS DTS platform, for detection of both CT and GC were compared.
METHODS
A total of 1018 first-void urine specimens were tested for the presence of CT and GC DNA using the 2 assays.
RESULTS
CT was detected in 143 specimens (14%). Eight specimens exhibited discordant results, and they were divided equally between the 2 assays. Based on the original results, the overall agreement for CT was 99.2%, with 97.1% and 99.5% in agreement with positive and negative specimens, respectively. Cohen's Kappa was 0.967. GC was detected in 27 specimens (2.6%). Two specimens exhibited discordant results, and they were divided equally between the 2 assays. Based on the original results, the overall agreement was 99.8%, with 96.2% and 99.9% in agreement for positive and negative specimens, respectively. Cohen's Kappa was 0.961.
CONCLUSIONS
There was a high level of agreement between the systems for both CT and GC detection.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.