Arm raising at exposure-controlled multidetector trauma CT of thoracoabdominal region: higher image quality, lower radiation dose.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Brink Monique, de Lange Frank, Oostveen Luuk J, Dekker Helena M, Kool Digna R, Deunk Jaap, Edwards Michael J R, van Kuijk Cornelis, Kamman Richard L, Blickman Johan G
Аннотация
PURPOSE: To evaluate the effect of arm position on image quality and effective radiation dose in an automatic exposure-controlled (AEC) multidetector thoracoabdominal computed tomography (CT) protocol in trauma patients.
MATERIALS AND METHODS: This retrospective study of the data of 177 trauma patients (117 male; median age, 39 years) was approved by the institutional ethics board, with informed patient consent waived. Patients underwent scanning by using an AEC 16-detector thoracoabdominal CT protocol in which both arms were raised above the shoulder region (standard-position group, 132 patients), one arm was raised and the other was down (one-arm group, 27 patients), or both arms were down (two-arm group, 18 patients). Objective and subjective image quality was assessed. Individual effective radiation dose was calculated from the effective tube current-time product per exposed section. For this purpose, section location-dependent conversion factors were derived by using a CT dosimetry calculator. The effect of arm position on effective dose was quantified by using linear regression analysis with correction for patient volume and attenuation.
RESULTS: Compared with the image quality in the standard-position group, the image quality in the one- and two-arm groups was decreased but within acceptable diagnostic limits. The median corrected effective dose in the standard-position group was 18.6 mSv; the dose in the one-arm group was 18% (95% confidence interval: 11%, 25%) higher than this, and that in the two-arm group was 45% (95% confidence interval: 34%, 57%) higher.
CONCLUSION: Omitting arm raising results in lower but acceptable image quality and a substantially higher effective radiation dose. Hence, effort should be made to position the arms above the shoulder when scanning trauma patients. Clinical trial registration no. NCT00228111.
MATERIALS AND METHODS: This retrospective study of the data of 177 trauma patients (117 male; median age, 39 years) was approved by the institutional ethics board, with informed patient consent waived. Patients underwent scanning by using an AEC 16-detector thoracoabdominal CT protocol in which both arms were raised above the shoulder region (standard-position group, 132 patients), one arm was raised and the other was down (one-arm group, 27 patients), or both arms were down (two-arm group, 18 patients). Objective and subjective image quality was assessed. Individual effective radiation dose was calculated from the effective tube current-time product per exposed section. For this purpose, section location-dependent conversion factors were derived by using a CT dosimetry calculator. The effect of arm position on effective dose was quantified by using linear regression analysis with correction for patient volume and attenuation.
RESULTS: Compared with the image quality in the standard-position group, the image quality in the one- and two-arm groups was decreased but within acceptable diagnostic limits. The median corrected effective dose in the standard-position group was 18.6 mSv; the dose in the one-arm group was 18% (95% confidence interval: 11%, 25%) higher than this, and that in the two-arm group was 45% (95% confidence interval: 34%, 57%) higher.
CONCLUSION: Omitting arm raising results in lower but acceptable image quality and a substantially higher effective radiation dose. Hence, effort should be made to position the arms above the shoulder when scanning trauma patients. Clinical trial registration no. NCT00228111.