m+Medelevator.RU
← Поиск публикаций
Journal of computer assisted tomography · 2015

Dual-Energy Computed Tomography-How Accurate Is Gemstone Spectrum Imaging Metal Artefact Reduction? Its Application to Orthopedic Metal Implants.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Dabirrahmani Danè, Magnussen John, Appleyard Richard Charles

Аннотация

OBJECTIVE: To assess the accuracy and suitability of dual-energy computed tomography (DECT) in scanning metals used in orthopedic implants.

MATERIALS AND METHODS: Four metal phantoms (Cobalt Chrome, Titanium Grade 5, Stainless Steel 316, and Stainless Steel 630), commonly used materials in orthopedic implants, were scanned by conventional, polychromatic CT as well as Gemstone Spectrum Imaging (GSI) DECT, with and without metal artefact reduction software (MARS). Scans were assessed for artefact based on Hounsfield unit values; and surfaces generated, based on a Canny edge detection algorithm. Two separate metal implants were also scanned and assessed for dimensional accuracy.

RESULTS: Conventional, polychromatic CT, and GSI DECT (without MARS) scans displayed major beam hardening in the presence of all four metals. The GSI DECT with MARS showed very clear and reproducible boundaries with minimal noise surrounding the metal phantoms. However, geometric analysis found overestimation of the dimensions, volume, and surface area for most of the metal phantoms. Titanium displayed the least artefact, compared to the other metals, in all scan scenarios.

CONCLUSIONS: Although metal artefact reduction using GSI DECT looks superior to conventional CT, when measured objectively, it was shown to overestimate geometries and skew dimensions. The GSI DECT with MARS should be used with caution, especially when assessing questions of implant shape or wear.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты