MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jacep.2023.09.004

Медицинская библиотека · JACC. Clinical electrophysiology · 2023

EsophAguS Deviation During RadiofrequencY Ablation of Atrial Fibrillation: The EASY AF Trial.

Weiss Raul, Osorio Jose, Nair Devi, Aguinaga Luis, Arabia Luis, Alcivar Diego, Al-Ahmad Amin, Tomassoni Gery, Kahaly Omar, Mehta Rohit, Ward Chad, Holmes Benjamin, Patel Dilesh, Killu Ammar M, Munger Thomas, Essandoh Michael, Houmsse Mahmoud, Rajendra Anil, Morales Gustavo, Hummel John D, Balasubramanian Gokulakrishan, Daoud Emile G

Рандомизированное исследованиеМногоцентровое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Injury to the esophagus has been reported in a high percentage of patients undergoing ablation of atrial fibrillation (AF).

OBJECTIVES

This study assessed the incidence of esophageal injury in patients undergoing ablation of AF with and without an esophageal deviating device.

METHODS

This prospective, randomized, multicenter, double-blinded, controlled Food and Drug Administration investigational device exemption trial compared the incidence of ablation-related esophageal lesions, as assessed by endoscopy, in patients undergoing AF ablation assigned to a control group (luminal esophageal temperature [LET] monitoring alone) compared with patients randomized to a deviation group (esophagus deviation device + LET). This novel deviating device uses vacuum suction and mechanical deflection to deviate a segment of the esophagus, including the trailing edge.

RESULTS

The data safety and monitoring board recommended stopping the study early after randomizing 120 patients due to deviating device efficacy. The primary study endpoint, ablation injury to the esophageal mucosa, was significantly less in the deviation group (5.7%) in comparison to the control group (35.4%; P < 0.0001). Control patients had a significantly higher severity and greater number of ablation lesions per patient. There was no adverse event assigned to the device. By multivariable analysis, the only feature associated with reduced esophageal lesions was randomization to deviating device (OR: 0.13; 95% CI: 0.04-0.46; P = 0.001). Among control subjects, there was no difference in esophageal lesions with high power/short duration (31.8%) vs other radiofrequency techniques (37.2%; P = 0.79).

CONCLUSIONS

The use of an esophageal deviating device resulted in a significant reduction in ablation-related esophageal lesions without any adverse events.

Для цитирования: Weiss Raul, Osorio Jose, Nair Devi, Aguinaga Luis, Arabia Luis, Alcivar Diego, Al-Ahmad Amin, Tomassoni Gery, Kahaly Omar, Mehta Rohit, Ward Chad, Holmes Benjamin, Patel Dilesh, Killu Ammar M, Munger Thomas, Essandoh Michael, Houmsse Mahmoud, Rajendra Anil, Morales Gustavo, Hummel John D, Balasubramanian Gokulakrishan, Daoud Emile G. EsophAguS Deviation During RadiofrequencY Ablation of Atrial Fibrillation: The EASY AF Trial.. JACC. Clinical electrophysiology. 2023. DOI: 10.1016/j.jacep.2023.09.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.