MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/s1743-9191(13)60014-x

Медицинская библиотека · International journal of surgery (London, England) · 2013

Continuous monitoring of the recurrent laryngeal nerve in thyroid surgery: a critical appraisal.

Dionigi Gianlorenzo, Donatini Gianluca, Boni Luigi, Rausei Stefano, Rovera Francesca, Tanda Maria Laura, Kim Hoon Yub, Chiang Feng-Yu, Wu Che-Wei, Mangano Alberto, Rulli Francesco, Alesina Piero F, Dionigi Renzo

СтатьяОбзорПеревод в очереди

Аннотация

BACKGROUND AND PURPOSE: Intraoperative neuromonitoring (IONM) contributes in several ways to recurrent laryngeal nerve (RLN) protection. Notwithstanding these advantages, surgeons must be aware that the current, intermittent, mode of IONM (I-IONM) has relevant limitations. To overcome these I-IONM limitations, a continuous IONM (C-IONM) technology has been proposed.

METHODS

A PubMed indexed literature review of the current limitations of I-IONM is presented and a commentary about C-IONM is provided presenting the preliminary results of research on this topic.

MAIN FINDINGS: I-IONM, despite the advantages it produces, presents some important limitations; to overcome these drawbacks a C-IONM technology has been introduced.

CONCLUSIONS

RLN traction injury is still the most common cause of RLN injury and is difficult to avoid with the application of I-IONM in thyroid surgery. C-IONM is useful to prevent the imminent traction injury by detecting progressive decreases in electromyographic amplitude combined with progressive latency increases. C-IONM seems to be a technological improvement. Likely, C-IONM by vagal nerve stimulation should enhance the standardization process, RLN intraoperative information, documentation, protection, training, and research in modern thyroid surgery. Although C-IONM is a promising technology at the cutting edge of research in thyroid surgery, we need more studies to assess in an evidence-based way all its advantages.

Для цитирования: Dionigi Gianlorenzo, Donatini Gianluca, Boni Luigi, Rausei Stefano, Rovera Francesca, Tanda Maria Laura, Kim Hoon Yub, Chiang Feng-Yu, Wu Che-Wei, Mangano Alberto, Rulli Francesco, Alesina Piero F, Dionigi Renzo. Continuous monitoring of the recurrent laryngeal nerve in thyroid surgery: a critical appraisal.. International journal of surgery (London, England). 2013. DOI: 10.1016/s1743-9191(13)60014-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.