Journal of orthopaedic surgery and research · 2017
Effect of different orientations of screw fixation for radial head fractures: a biomechanical comparison.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Shi Xuchao, Pan Tianlong, Wu Dengying, Cai Ningyu, Chen Rong, Li Bin, Zhang Rui, Zhou Chengwei, Pan Jun
Аннотация
BACKGROUND: Screw fixation is a common method used for the treatment of Mason type II radial head fractures. The purpose of our study was to evaluate the mechanical properties of three different screw orientations used for fixation of Mason type II radial head fractures.
METHODS: We sawed 24 medium-frequency fourth-generation Synbone radial bones to simulate unstable radial head fractures, which we then fixed with three different screw orientations. Implants were tested under axial load by the tension-torsion composite test system. If the implant-radial constructs did not fail after the axial load test, an axial failure load was added to the remaining constructs.
RESULTS: The stiffness of the divergent group was the highest of the three orientations, and this group had statistically significant difference from the other two groups (p < 0.05). However, there was no statistically significant difference between the convergence group and the parallel group (p > 0.05). When the displacement reached 2 mm, the load of the divergent screw was still larger than the other two groups (p < 0.05).
CONCLUSIONS: The divergent screw orientation was the most stable and had the greatest control of Mason type II fractures of these three groups. Therefore, it can be better applied in clinical settings.
METHODS: We sawed 24 medium-frequency fourth-generation Synbone radial bones to simulate unstable radial head fractures, which we then fixed with three different screw orientations. Implants were tested under axial load by the tension-torsion composite test system. If the implant-radial constructs did not fail after the axial load test, an axial failure load was added to the remaining constructs.
RESULTS: The stiffness of the divergent group was the highest of the three orientations, and this group had statistically significant difference from the other two groups (p < 0.05). However, there was no statistically significant difference between the convergence group and the parallel group (p > 0.05). When the displacement reached 2 mm, the load of the divergent screw was still larger than the other two groups (p < 0.05).
CONCLUSIONS: The divergent screw orientation was the most stable and had the greatest control of Mason type II fractures of these three groups. Therefore, it can be better applied in clinical settings.