Use of portable single-lead electrocardiogram device as an alternative for QTc monitoring in critically ill patients.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Rebolledo-Del Toro Martin, Carvajalino-Galeano Ana Beatriz, Pinto-Brito Clarena, Muñoz-Velandia Oscar Mauricio, García-Peña Ángel Alberto
Аннотация
PURPOSE: Acquired QT prolongation is frequent and leads to a higher mortality rate in critically ill patients. KardiaMobile 1L® (KM1L) is a portable, user-friendly single lead, mobile alternative to conventional 12-lead electrocardiogram (12-L ECG) that could be more readily available, potentially facilitating more frequent QTc assessments in intensive care units (ICU); however, there is currently no evidence to validate this potential use.
METHODS: We conducted a prospective diagnostic test study comparing QT interval measurement using KM1L with conventional 12-L ECG ordered for any reason in patients admitted to an ICU. We compared the mean difference using a paired t-test, agreement using Bland-Altman analysis, and Lin's concordance coefficient, numerical precision (proportion of QT measurements with <10 ms difference between KM1L and conventional 12-L ECG), and clinical precision (concordance for adequate discrimination of prolonged QTc).
RESULTS: We included 114 patients (61.4% men, 60% cardiovascular etiology of hospitalization) with 131 12-L ECG traces. We found no statistical difference between corrected QT measurements (427 ms vs. 428 ms, p = .308). Lin's concordance coefficient was 0.848 (95% CI 0.801-0.894, p = .001). Clinical precision was excellent in males and substantial in females (Kappa 0.837 and 0.781, respectively). Numerical precision was lower in patients with vasoactive drugs (-13.99 ms), QT-prolonging drugs (13.84 ms), antiarrhythmic drugs (-12.87 ms), and a heart rate (HR) difference of ≥5 beats per minute (bpm) between devices (-11.26 ms).
CONCLUSION: Our study validates the clinical viability of KM1L, a single-lead mobile ECG device, for identifying prolonged QT intervals in ICU patients. Caution is warranted in patients with certain medical conditions that may affect numerical precision.
METHODS: We conducted a prospective diagnostic test study comparing QT interval measurement using KM1L with conventional 12-L ECG ordered for any reason in patients admitted to an ICU. We compared the mean difference using a paired t-test, agreement using Bland-Altman analysis, and Lin's concordance coefficient, numerical precision (proportion of QT measurements with <10 ms difference between KM1L and conventional 12-L ECG), and clinical precision (concordance for adequate discrimination of prolonged QTc).
RESULTS: We included 114 patients (61.4% men, 60% cardiovascular etiology of hospitalization) with 131 12-L ECG traces. We found no statistical difference between corrected QT measurements (427 ms vs. 428 ms, p = .308). Lin's concordance coefficient was 0.848 (95% CI 0.801-0.894, p = .001). Clinical precision was excellent in males and substantial in females (Kappa 0.837 and 0.781, respectively). Numerical precision was lower in patients with vasoactive drugs (-13.99 ms), QT-prolonging drugs (13.84 ms), antiarrhythmic drugs (-12.87 ms), and a heart rate (HR) difference of ≥5 beats per minute (bpm) between devices (-11.26 ms).
CONCLUSION: Our study validates the clinical viability of KM1L, a single-lead mobile ECG device, for identifying prolonged QT intervals in ICU patients. Caution is warranted in patients with certain medical conditions that may affect numerical precision.