m+Medelevator.RU
← Поиск публикаций
Journal of cardiology · 2020

Percutaneous pulmonary artery debanding.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Tomita Hideshi, Fujii Takanari, Kise Hiroaki, Oyama Nobuo, Shimizu Takeshi, Nagaoka Kota, Tarui Suguru, Miyahara Yoshinori, Ishino Kozo

Аннотация

BACKGROUND: There is a paucity of data on palliative or total percutaneous pulmonary artery debanding (p-debanding), particularly with use of a stent.

METHODS: Twelve p-debandings in eight patients were included in this study. Age at pulmonary artery banding (PAB) ranged from 3 days to 1 year (median, 13 days), while p-debanding was performed at 2-157 (7) months. The body weight at the p-debanding ranged from 3.2 to 22.2 (7.3) kg. We chose the balloon diameter of 30-50% to the circumference of the band for palliative, and larger than 50% for total p-debanding, respectively. In either way, the balloon diameter did not exceed 1.5 times the reference vessel diameter. Stent was implanted for palliative p-debanding in 2 patients.

RESULTS: 1. The circumference of the band ranged from 16 to 23 (20) mm, while the balloon diameter ranged from 20-60 (40)% to that, where larger than 50% was used for 2 procedures intended total p-debanding. 2. PAB diameter increased from 2.5-4.7 (3.0) mm to 2.8-9.5 (4.5) mm (p<0.01), however, there was no significant change in the diameter in 2 procedures. In one patient, p-debanding was the definitive treatment associated with spontaneous near closure of muscular ventricular septal defect, in another patient of congenitally corrected transposition of the great arteries, severely depressed left ventricular ejection fraction was recovered following p-debanding. 3. Arterial oxygen saturation (SaO2) increased from 64-97 (80)% to 66-95 (90)% (p<0.01), while in 10 procedures of 6 patients where the indication of p-debanding was hypoxia, SaO2 increased in 8 procedures. There was no significant pulmonary hypertension following p-debanding.

CONCLUSION: Palliative or total p-debanding using balloon and/or stenting is generally feasible and effective. A balloon diameter 35-50% to the band circumference in palliative, and more than 50% in total p-debanding, while in either way less than 1.5 times the reference vessel diameter, is safe.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты