Risk of extravasation after power injection of contrast media via the proximal port of multilumen central venous catheters: case report and review of the literature.
СтатьяОбзорНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Schummer C, Sakr Y, Steenbeck J, Gugel M, Reinhart K, Schummer W
Аннотация
PURPOSE: Multilumen central venous catheters (CVCs) are not commonly used for power injection. However, in critically ill patients, CVCs-- most of which do not have FDA approval for power injection--may be the only available venous access.
MATERIALS AND METHODS: The pitfalls of multilumen CVCs are illustrated by a case report of a patient in whom extravasation of intravenously administered contrast medium occurred after power injection in a triple-lumen CVC using the lumen with the port furthest from the catheter tip.
RESULTS: The underlying mechanisms for the displacement of the initially correctly placed right subclavian CVC could include elevation of both arms of the obese patient or the power injection itself. The distances between port openings and catheter tips of various commercially available multilumen CVCs are assessed. We examine the possible caveats of ECG-guided CVC placement for optimal tip position, discuss technical difficulties related to power injection via CVCs, and review commonly used drugs that may cause extravasation injury.
CONCLUSION: Knowledge of the distances between CVC port openings and the catheter tip are essential for safe intravasal administration of fluids.
MATERIALS AND METHODS: The pitfalls of multilumen CVCs are illustrated by a case report of a patient in whom extravasation of intravenously administered contrast medium occurred after power injection in a triple-lumen CVC using the lumen with the port furthest from the catheter tip.
RESULTS: The underlying mechanisms for the displacement of the initially correctly placed right subclavian CVC could include elevation of both arms of the obese patient or the power injection itself. The distances between port openings and catheter tips of various commercially available multilumen CVCs are assessed. We examine the possible caveats of ECG-guided CVC placement for optimal tip position, discuss technical difficulties related to power injection via CVCs, and review commonly used drugs that may cause extravasation injury.
CONCLUSION: Knowledge of the distances between CVC port openings and the catheter tip are essential for safe intravasal administration of fluids.
Предметные рубрики MeSH
AdultCatheterization, Central VenousContrast MediaExtravasation of Diagnostic and Therapeutic MaterialsFemaleHumansInjections, IntravenousIntensive Care UnitsIohexolLiver AbscessLung DiseasesMediastinumNeckObesityQuality Assurance, Health CareShock, SepticShoulderSubclavian VeinTomography, Spiral ComputedVena Cava, Superior