Tabrizi Sepehr N, Chen Shujun, Tapsall John, Garland Suzanne M
Аннотация
OBJECTIVES: Detection of Neisseria gonorrhoeae by commercial and in-house-based assays has been hampered by false-positive and false-negative results. The current study describes a sensitive and specific real-time 5'-nuclease PCR assay targeting a 90-bp region of the multicopy opa gene.
GOAL: To evaluate the sensitivity and specificity of this assay in detection of gonococcus.
STUDY: Sensitivity and specificity were determined by testing a panel of 173 microorganisms. In addition, 135 clinical samples previously evaluated by 4 nucleic acid amplification methods were also tested.
RESULTS: A sensitivity of 1 copy per reaction was achieved. Positive results were only obtained for N gonorrhoeae strains including 20 cppB-negative strains. Overall, 134 of 135 clinical sample results agreed with the consensus nucleic amplification methods.
CONCLUSION: This study demonstrates opa-based target can be used as an accurate and rapid PCR assay for the detection of N gonorrhoeae in clinical specimens.
GOAL: To evaluate the sensitivity and specificity of this assay in detection of gonococcus.
STUDY: Sensitivity and specificity were determined by testing a panel of 173 microorganisms. In addition, 135 clinical samples previously evaluated by 4 nucleic acid amplification methods were also tested.
RESULTS: A sensitivity of 1 copy per reaction was achieved. Positive results were only obtained for N gonorrhoeae strains including 20 cppB-negative strains. Overall, 134 of 135 clinical sample results agreed with the consensus nucleic amplification methods.
CONCLUSION: This study demonstrates opa-based target can be used as an accurate and rapid PCR assay for the detection of N gonorrhoeae in clinical specimens.